STD X-ray Kodak 2000 fix error E13
Profile:Fixing error E13 in dental X-ray Kodak 2000 (2100, 2200) when this is caused by a breaking in the lower resistor of the high voltage divider.
License: Public Domain
Dental X-ray Kodak 2000, sometimes has a problem with the internal voltage divider of kV for feedback in the HV tank. Interrupts the bottom resistor of the high voltage divider.
The easy and correct solution is new tube tank.
less correct solution is disassembly of the tube and soldering of the resistor.
Extremal Solition this problem is here: https://www.ebme.co.uk/forums/ubbthreads.php/ubb/printthread/Board/2/main/9999/type/thread
This circuit has not been tested by my. Everything you do, you do at your own risk! I am not responsible for any damage!
PCB_X-ray Kodak 2000 fix error E13
Open in EditorPCB_X-ray Kodak 2000 fix error E13 v2
Open in EditorID | Name | Designator | Footprint | Quantity |
---|---|---|---|---|
1 | 1n | C1 | RAD-0.2 | 1 |
2 | 10n | C2 | RAD-0.2 | 1 |
3 | 1u | C3,C4 | CAP-D8.0×F3.5 | 2 |
4 | 1N4148 | D1,D2 | DO-35_BD2.0-L4.2-P8.20-D0.5-RD | 2 |
5 | HDR-F-2.54_1x4 | H1 | HDR-F-2.54_1X4 | 1 |
6 | 10k | R1,R2,R3 | R_AXIAL-0.4 | 3 |
7 | 13k | R4 | R_AXIAL-0.4 | 1 |
8 | 1 | R5 | R_AXIAL-0.4 | 1 |
9 | 10 | R6 | R_AXIAL-0.4 | 1 |
10 | TL082AD | U1 | PDIP-8_L10.2-W6.5-P2.54-LS7.6-BL | 1 |