OngoingPeloton Medical Prototype 2
STDPeloton Medical Prototype 2
License
:Public Domain
Description
Once the reed switch is activated, the first LM555 timer acts as a delay (to delay the activation of the motor) before the second LM555 timer is activated. The purpose of this delay (the first LM555) is to prevent any unintentional reed switch triggers to trigger the motor, thus if the reed switch is shut off before the first timer completes, the motor won't activate. If the first timer completes, the second timer is activated and turns the motor on for a set amount of time which will result in basically a 180-degree rotation of the motor and will clamp the venous tubing.
Design Drawing
The preview image was not generated, please save it again in the editor.BOM
Bom empty
CloneProject Members
Intellectual Property Statement & Reproduction Instructions
This is an open-source hardware project. All intellectual property rights belong to the creator. The project is shared on the platform for learning, communication, and research only; any commercial use is prohibited. If your intellectual property rights are infringed on EasyEDA, please notify us by submitting relevant materials in accordance with the Rules for Complaints and Appeals of IPR Infringement.